Calibrated Light Source

Calibrated Light SourceSpectrum correction for Ex side can be automatically performed by placing the Rhodamine B solution in front of the detector of reference side. Spectrum correction for Em side is performed by utilizing the correction factor which is based on corrected Ex spectrum. But, this procedure enables to correct the wavelength range from 220 nm to 600 nm because of using Rhodamine B solution. In order to correct the longer wavelength range from 600 nm, using another calibrated standard light source is required. Combining the procedure Rhodamine B solution with that of standard light source is capable of correcting the full wavelength range. This is a calibrated tungsten lamp for spectrum correction over 600 nm. A calibrated light source for the FP-6200 is also available as customized order.
For FP-6600/6500/6300
6808-J007A ESC-333 Calibrated light source |