6855-J032A micro21AS-01 Autosampler
|
| Sample injection method : |
|
Variable sample volume injection method Time injection, Loop injection (Flow line changing is required.) |
| Metering method : |
|
By metering pump |
| Metering pump speed : |
|
0.1 ~ 25.0 μL/sec
(standard metering pump) variable setting |
| Injection volume: |
|
0.1 ~ 100 μL (0.1 μL increments) |
| Number of samples : |
|
120 (2.0 mL vials, standard rack)
224 (300 μL or 200 μL vials, micro-volume rack)
96 x 2 (96-well microplate X 2, microplate rack)
384 x 2 (384-well microplate X 2, microplate rack) |
| Reproducibility: |
|
RSD 0.3 % or less (10 ~ 100 μL),
RSD 1.0 % or less (1.1 ~ 9.9 μL),
RSD 5.0 % or less (1.0 μL or less)
(Using designated parameters under specified conditions)
(When Loop injection method is selected, less than 1.0 % RSD at 1.0 μL injection using designated parameters under specified conditions.) |
| Contamination : |
|
0.01 % or less for 10 μL injection using designated parameters under specified conditions |
| Dimension: |
|
300 (W) x 386 (H) x 500 (D) mm |
| Weight: |
|
Approx. 25 kg |
| Power Input voltage: |
|
AC100 ~ 240 V ±10% 50/60 Hz |
| Power consumption: |
|
230 VA |