| 6855-J031A micro21LC-01 Capillary Liquid Chromatograph (manual injection) |
6855-J041A micro21LC-02 Capillary Liquid Chromatograph (auto injection)
|
Pump unit |
| Flow rate setting : |
|
0.1 ~ 300 μL/min with 0.1 μL increments |
| Recommended flow rate range : |
|
Isocratic 0.2 ~ 200 μL/min |
|
|
Gradient 1.0 ~ 200 μL/min |
| Flow rate accuracy : |
|
±4% of setting or ±0.08 μL/min whichever is greater at 1.0 ~ 200 μL/min, 1 ~ 25 MPa |
| Flow rate precision : |
|
Less than 0.8 % (Relative Standard Deviation) at 1.0 ~ 200 μL/min |
| Max. pressure : |
|
35 MPa (less than 100 μL/min)
|
|
|
25 MPa (More than 100 μL/min)
|
| Pressure limit : |
|
0 ~ 35 MPa, Settable max. and min. pressure limit in 0.1 MPa increments
|
| Gradient composition range : |
|
0 ~ 100 % with 0.1 % increments
|
| Gradient precision : |
|
±3.0 %
|
|
|
|
Degasser part |
| Number of lines: |
|
4 |
| Solvent hold-up volume: |
|
Approx. 145 μL each line
|
| Maximum flowrate: |
|
200 μL/min |
| Maximum Pressure: |
|
0.2 MPa |
| Mixing part |
|
|
| Mixing method : |
|
By rotation of stirrer bar |
| Number of inlet ports : |
|
2 |
| Mixing chamber capacity : |
|
20 μL standard, 47, 80 μL optional |
|
|
|
Column Oven part |
| Temperature control range: |
|
15°C below ambient temperature to 65°C |
| Temperature setting range: |
|
4°C ~ 65°C (set in increments of 0.1°C) |
| Column compartment: |
|
320(W) x 20(H) x 18(D) mm |
|
|
|
Others |
| Dimension: |
|
450 (W) x 300 (H) x 500 (D) mm |
| Weight: |
|
Approx. 38 kg |
| Power Input voltage: |
|
AC100 ~ 240 V ±10% 50/60 Hz |
| Power consumption: |
|
150 VA |